Analytical validation of a β‑hydroxybutyrate assay on the BT3500 analyzer for feline serum
Accurate measurement of serum β‑hydroxybutyrate is critical for managing diabetic cats, especially when using SGLT2 inhibitors. This study evaluated the Randox Ranbut enzymatic assay on the BT3500 chemistry analyzer, demonstrating acceptable precision, linearity, and sensitivity, and defining short‑ and long‑term storage stability in feline serum.
Study and findings
The investigators used residual feline serum and pooled samples to assess analytical performance of the Randox β‑OHB assay on the BT3500 platform. Intra‑assay coefficient of variation ranged from 1.68 % to 14.48 %, with higher variability at the lowest concentrations; inter‑assay CV ranged from 3.08 % to 10.20 %. Spike‑recovery experiments across five concentration levels yielded recoveries between 110 % and 122 % (mean 114 % ± 4 %). Serial dilutions produced a linear response (slope = 1.012, R² ≥ 0.9998). The limit of blank, detection, and quantitation were 0.044, 0.091, and 0.150 mmol/L, respectively. Stability testing showed that β‑OHB remained unchanged for up to six days at 4 °C and for up to 180 days at –20 °C; a measurable decline appeared after 360 days of frozen storage.
Clinical interpretation
These analytical characteristics indicate that the BT3500, when coupled with the Randox enzymatic kit, can reliably quantify β‑OHB in feline serum within the clinically relevant range for detecting ketosis and ketoacidosis. The modest imprecision at low concentrations should be considered when interpreting marginally elevated values, but the overall accuracy and linearity support routine use in veterinary practice. The defined storage limits provide practical guidance: serum can be refrigerated for a week without loss of signal, and frozen samples remain trustworthy for up to six months, facilitating batch testing or retrospective analyses.
Limitations and open questions
The validation was performed on residual and pooled samples rather than on a prospectively recruited cohort of diabetic cats, so the study does not address biological variability or clinical decision thresholds in a real‑world population. Additionally, the impact of hemolysis, lipemia, or other common serum interferences was not examined. Future work should evaluate the assay’s performance across a broader spectrum of disease states and assess whether the observed analytical bias at low concentrations influences clinical classification of euglycemic ketoacidosis.
Source
Tirolo A et al. Analytical Validation of β‑Hydroxybutyrate Quantification in Feline Serum with the BT3500 Analyzer. Veterinary Clinical Pathology. 2026.